Development of probe electrospray using a solid needle
Development of probe electrospray using a solid needle
复制标题
DOI:
10.1002/rcm.3201
复制
发表时间:
2007-01-01
影响因子:
2
通讯作者:
Suzuki, Shigeo
中科院分区:
文献类型:
--
作者:
Hiraoka, Kenzo;Nishidate, Kentaro;Suzuki, Shigeo
Electrospray using a solid needle probe for the analysis of liquid samples was developed. The needle was moved up and down a vertical axis by a motor-driven system. At the lowest position of the needle, the tip of the needle touched the surface of the liquid sample and a small amount of liquid was picked up by the needle. At the highest position, a high voltage of similar to 3 kV was applied to the needle and the sample loaded on the tip of the needle was electrosprayed. The ions formed were measured by an orthogonal time-of-flight mass spectrometer. A single sample loading by the needle gave reasonably strong ion signals for amino acids, peptides, proteins and polyethylene glycol (PEG) in aqueous solution. The addition of salts or acids to aqueous solution led to dramatic enhancement of the ion signals. Copyright (c) 2007 John Wiley & Sons, Ltd.