Determination of psilocin, bufotenine, LSD and its metabolites in serum, plasma and urine by SPE-LC-MS/MS
Determination of psilocin, bufotenine, LSD and its metabolites in serum, plasma and urine by SPE-LC-MS/MS
复制标题
DOI:
10.1007/s00414-012-0796-1
复制
发表时间:
2013-05-01
影响因子:
2.1
通讯作者:
Koehler, Helga
中科院分区:
文献类型:
--
作者:
Martin, Rafaela;Schuerenkamp, Jennifer;Koehler, Helga
A validated method for the simultaneous determination of psilocin, bufotenine, lysergic acid diethylamide and its metabolites in serum, plasma and urine using liquid chromatography-electrospray ionization/tandem mass spectrometry was developed. During the solid-phase extraction procedure with polymeric mixed-mode cation exchange columns, the unstable analytes were protected by ascorbic acid, drying with nitrogen and exclusion of light. The limits of detection and quantitation for all analytes were low. Recovery was a parts per thousand yen86 % for all analytes and no significant matrix effects were observed. Interday and intraday imprecisions at different concentrations ranged from 1.1 to 8.2 % relative standard deviation, bias was within +/- 5.3 %. Processed samples were stable in the autosampler for at least 2 days. Furthermore, freeze/thaw and long-term stability were investigated. The method was successfully applied to authentic serum and urine samples.