A novel device (SD-101) with high accuracy for screening sleep apnoea-hypopnoea syndrome

A novel device (SD-101) with high accuracy for screening sleep apnoea-hypopnoea syndrome
复制标题

DOI:
10.1111/j.1440-1843.2009.01627.x
复制
发表时间:
2009-11-01
期刊:
影响因子:
6.9
通讯作者:
Nomiyama, Tetsuo
Nomiyama, Tetsuo
中科院分区:
医学2区
文献类型:
--
作者:
Agatsuma, Toshihiko;Fujimoto, Keisaku;Nomiyama, Tetsuo

文献摘要

被引文献

相似文献

背景和目的:SD-101是一种非限制性的片状医疗设备,带有一系列压力传感器,通过感测与呼吸运动相对应的身体重力变化来检测睡眠呼吸障碍。为评价SD-101与多导睡眠图(PSG)筛查睡眠呼吸暂停低通气综合征(SAHS)的准确性,对201例疑似SAHS患者(SAHS组)和165例某运输公司男性职工(筛查组)同时进行了夜间多导睡眠图和SD-101监测。结果:多导睡眠图显示AHI值为1.54 ± 0.001,AHI值为0.001 ± 0.001,两组比较差异有统计学意义(P < 0.01)。结论:SD-101是一种有效的SAHS筛查方法,可用于SAHS的筛查
Background and objective: The SD-101 is a non-restrictive, sheet-like medical device with an array of pressure sensors, to detect sleep-disordered breathing by sensing gravitational alterations in the body corresponding to respiratory movements. This study evaluated the accuracy of the SD-101 for screening sleep apnoea-hypopnoea syndrome (SAHS) by comparison with polysomnography.Methods: Nocturnal polysomnography and SD-101 monitoring were conducted simultaneously and compared in 201 patients with suspected SAHS (suspected SAHS group) and 165 male employees of a transport company (screening group).Results: Polysomnography revealed an AHI of