Biophysical characteristics of HIMAC clinical irradiation system for heavy-ion radiation therapy
Biophysical characteristics of HIMAC clinical irradiation system for heavy-ion radiation therapy
复制标题
DOI:
10.1016/s0360-3016(98)00544-6
复制
发表时间:
1999-04-01
影响因子:
7
通讯作者:
Kawachi, K
中科院分区:
文献类型:
--
作者:
Kanai, T;Endo, M;Kawachi, K
Purpose: The irradiation system and biophysical characteristic carbon beams are examined regarding radiation therapy.Methods and Materials: An irradiation system was developed for heavy-ion radiotherapy. Wobbler magnets and a scatterer were used for flattening the radiation field. A patient-positioning system using X ray and image intensifiers was also installed in the irradiation system. The depth-dose distributions of the carbon beams were modified to make a spread-out Bragg peak, which was designed based on the biophysical characteristics of monoenergetic beams. A dosimetry system for heavy-ion radiotherapy was established to deliver heavy-ion doses safely to the patients according to the treatment planning. A carbon beam of 80 keV/mu m in the spread-out Bragg peak was found to be equivalent in biological responses to the neutron beam that is produced at cyclotron facility in National Institute Radiological Sciences (NIRS) by bombarding 30-MeV deuteron beam on beryllium target. The fractionation schedule of the NIRS neutron therapy was adapted for the first clinical trials using carbon beams.Results: Carbon beams, 290, 350, and 400 MeV/u, were used for a clinical trial from June of 1994. Over 300 patients have already been treated by this irradiation system by the end of 1997. (C) 1999 Elsevier Science Inc.