Ultra-performance liquid chromatography/mass spectrometry of intact proteins
Ultra-performance liquid chromatography/mass spectrometry of intact proteins
复制标题
DOI:
10.1016/j.chroma.2008.03.058
复制
发表时间:
2008-05-30
影响因子:
4.1
通讯作者:
Croley, Timothy R.
中科院分区:
文献类型:
--
作者:
Everley, Robert A.;Croley, Timothy R.
Given that numerous small molecule applications of ultra-performance liquid chromatography (UPLC) have been published, efforts were made to examine the potential of UPLC to enhance the separation of intact proteins. Beginning with typically employed conditions, column temperature and organic solvent were optimized followed by an HPLC vs. UPLC comparison. When applied to a mixture of 10 protein standards, the optimized method yielded improved chromatographic resolution, enhanced sensitivity, and a threefold increase in throughput. Subsequent cell lysate analysis demonstrated no compromise in chromatographic or mass spectral data quality at 1/3 of the original run time. (c) 2008 Elsevier B.V. All rights reserved.