A fixed frequency aerosol albedometer.
A fixed frequency aerosol albedometer.
复制标题
DOI:
10.1364/oe.16.002191
复制
发表时间:
2008-02
期刊:
影响因子:
3.8
通讯作者:
Jonathan E. Thompson;Nick Barta;Danielle M. Policarpio;Richard Duvall
中科院分区:
文献类型:
--
作者:
Jonathan E. Thompson;Nick Barta;Danielle M. Policarpio;Richard Duvall
A new method for the measurement of aerosol single scatter albedo (omega) at 532 nm was developed. The method employs cavity ring-down spectroscopy (CRDS) for measurement of aerosol extinction coefficient (b(ext)) and an integrating sphere nephelometer for determination of aerosol scattering coefficient (b(scat)). A unique feature of this method is that the extinction and scattering measurements are conducted simultaneously, on the exact same sample volume. Limits of detection (3s) for the extinction and scattering channel were 0.61 Mm(-1) and 2.7 Mm(-1) respectively.