Anterior Ocular Biometry Using 3-Dimensional Optical Coherence Tomography
Anterior Ocular Biometry Using 3-Dimensional Optical Coherence Tomography
复制标题
DOI:
10.1016/j.ophtha.2008.12.022
复制
发表时间:
2009-05-01
期刊:
影响因子:
13.7
通讯作者:
Oshika, Tetsuro
中科院分区:
文献类型:
--
作者:
Fukuda, Shinichi;Kawana, Keisuke;Oshika, Tetsuro
Purpose: To evaluate anterior ocular biometry by comparing the measurements of central corneal thickness (CCT) and anterior chamber depth (ACD) with 3-dimensional corneal and anterior segment optical coherence tomography (CAS-OCT) and other methods.Design: Cross-sectional study.Participants: Forty eyes of 40 normal subjects.Methods: The CCT was measured by 4 methods (CAS-OCT, Scheimpflug camera, scanning-slit topography, and ultrasonic pachymetry), and the ACD was measured by 3 methods (CAS-OCT, Scheimpflug camera, and scanning-slit topography). The anterior chamber volume (ACV) was calculated with CAS-OCT. Repeatability and reproducibility of CAS-OCT measurements were evaluated.Main Outcome Measurements: The CCT and ACD were measured and compared between devices. The ACV was calculated with CAS-OCT. Coefficient of variation and intraclass correlation coefficient (ICC) were calculated to evaluate repeatability and reproducibility of CAS-OCT measurements.Results: The mean CCT was 547.0 +/- 39.0, 590.0 +/- 39.8, 525.0 +/- 45.0, and 545.0 +/- 40.3 mu m with CAS-OCT, Scheimpflug camera, scanning-slit topography, and ultrasonic pachymetry, respectively. Significant differences were observed among the 4 methods (P