Reliability of the electronic early treatment diabetic retinopathy study testing protocol in children 7 to <13 years old

Reliability of the electronic early treatment diabetic retinopathy study testing protocol in children 7 to <13 years old
复制标题

DOI:
10.1016/s0002-9394(03)00388-x
复制
发表时间:
2003-10-01
影响因子:
4.2
通讯作者:
Moke, PS
Moke, PS
中科院分区:
医学1区
文献类型:
--
作者:
Cotter, SA;Chu, RH;Moke, PS

文献摘要

被引文献

相似文献

目的:利用计算机化电子视力(EVA)测试仪对7 ~ 13岁以下儿童进行电子早期治疗糖尿病视网膜病变研究(E-ETDRS)视力算法的重测信度评估。设计:重测信度研究。方法:这项多中心研究涉及四个临床站点的245名受试者。作为主要结果测量,使用EVA系统上的E-ETDRS测试协议对视力进行了两次测量,该系统使用可编程手持设备与个人计算机和17英寸显示器在3米测试距离上进行通信。结果:重测信度高(右眼r = 0.94,左眼r = 0.96),左眼和右眼的重测分数89%在初始测试分数的最小分辨角(logMAR)(5个字母)的0.1对数范围内,99%在0.2 logMAR(10个字母)范围内。信度在7岁至7岁年龄组中较高
PURPOSE: To assess the test-retest reliability of the electronic Early Treatment Diabetic Retinopathy Study (E-ETDRS) visual acuity algorithm using the computerized Electronic Visual Acuity (EVA) tester in children 7 to < 13 years old.DESIGN: Test-retest reliability study.METHODS: This multicenter study involved 245 subjects at four clinical sites. As the main outcome measure, visual acuity was measured twice using the E-ETDRS testing protocol on the EVA system, which uses a programmed handheld device to communicate with a personal computer and a 17,inch monitor at a 3-m test distance.RESULTS: Test-retest reliability was high (r = .94 for right eyes and 0.96 for left eyes) and for both right and left eyes, 89% of retest scores were within 0.1 logarithm of the minimal angle of resolution (logMAR) (five let, ters) of the initial test score and 99% of retests were within 0.2 logMAR (10 letters). Reliability was high across the age range of 7 to