Filtered back projection revisited in low-kilovolt computed tomography angiography: sharp filter kernel enhances visualization of the artery of Adamkiewicz
Filtered back projection revisited in low-kilovolt computed tomography angiography: sharp filter kernel enhances visualization of the artery of Adamkiewicz
复制标题
DOI:
10.1007/s00234-018-2136-8
复制
发表时间:
2019-03-01
期刊:
影响因子:
2.8
通讯作者:
Fukuda, Tetsuya
中科院分区:
文献类型:
--
作者:
Nishii, Tatsuya;Kotoku, Akiyuki;Fukuda, Tetsuya
PurposeThe reliability of assessment of the artery of Adamkiewicz before the aortic repair is highly dependent on the display of the continuity of this artery with the aorta, mainly around the vertebral pedicle, by computed tomography angiography (CTA). We hypothesized that the sharp filter kernel can improve visualization of this continuity of the vessel structure because of its edge enhancement and high-spatial resolution. This study was performed to compare the subjective and objective image quality of spinal CTA reconstructed with sharp and smooth filter kernels.MethodsWe retrospectively reviewed 40 consecutive patients who had undergone 80-kV CTA to detect the artery of Adamkiewicz before aortic repair. We measured the CT number and the contrast-to-noise ratio of the anterior spinal artery to the spinal cord. Furthermore, the continuity of the artery of Adamkiewicz was evaluated using a 3-point scale (2 points, absolute; 0 points, undetectable).ResultsCTA with the sharp filter kernel showed a significantly higher CT number and contrast-to-noise ratio of the spinal artery than did CTA with the smooth filter kernel (P