A Wearable Device for Minimally-Invasive Behind-the-Ear EEG and Evoked Potentials
A Wearable Device for Minimally-Invasive Behind-the-Ear EEG and Evoked Potentials
复制标题
用于微创耳后脑电图和诱发电位的可穿戴设备
DOI:
10.1109/biocas.2018.8584814
复制
发表时间:
2018
期刊:
影响因子:
--
通讯作者:
Luca Bertini
中科院分区:
文献类型:
--
作者:
M. Guermandi;Simone Benatti;Victor Javier Kartsch Morinigo;Luca Bertini
We present an unobtrusive and minimally invasive system for acquiring 8-channels of EEG behind the ear, to support brain activity monitoring in a socially acceptable fashion and outside clinical environments. Electrical performance are in line with those of clinical EEG systems (0.48 µVRMS integrated noise in the [0-100] Hz band, 106 dB CMRR) and setup is extremely fast, requiring only a mild cleaning of the skin before applying the device. The system can acquire typical brain activity rhythms and potentials evoked by sensory stimuli, which are at the basis of BCI systems. Current consumption is 6.2 rnA, plus 16 mA for Bluetooth® data transmission (4.5 hours life on a 100 mAh battery). The overall cost of components is below 50 USD.