Precise measurements of trace neodymium isotopes as Nd ions by thermal ionization mass spectrometry using film porous ion emitters
Precise measurements of trace neodymium isotopes as Nd ions by thermal ionization mass spectrometry using film porous ion emitters
复制标题
使用薄膜多孔离子发射器通过热电离质谱法精确测量 Nd 离子形式的痕量钕同位素
DOI:
10.1016/j.sab.2019.105656
复制
发表时间:
2019
期刊:
影响因子:
--
通讯作者:
Hao Fanhua
中科院分区:
文献类型:
--
作者:
Shao Xuepeng;Bu Wenting;Long Kaiming;Tang Lei;Liu Xuemei;Yan Heng;Hao Fanhua
Neodymium isotope analysis as Nd+ions is more favored than as NdO+ions due to less potential isobaric interferences and no need for complicated oxygen isotope corrections. However, the poor sensitivity and inadequate reproducibility seriously hinder its application. In this study, a new analytical method by thermal ionization mass spectrometry (TIMS) is initially presented for the precise measurement of143Nd/144Nd ratio as Nd+ions with film porous ion emitter (FPIE), combined with a simplified Nd chemical separation procedure. The Nd+ion yields were one order of magnitude higher than those of traditional Nd+ion analysis method. Repeated measurements of reference material JNdi-1 (1 ng) yielded a143Nd/144Nd value of 0.512113 ± 33 (2SD,n= 9). The achieved external precisions fulfilled the requirements for applications in nuclear forensics. Rock powder reference materials and uranium ore samples were analyzed to further test the accuracy and reproducibility of our method. We believe that this method is not only a particular value for nuclear forensic purposes, but also can expand the applications of Nd isotope ratio analysis in geochemistry, geochronology and environmental sciences.