WINDOWLESS, FLOW TYPE, PROPORTIONAL COUNTER FOR COUNTING C-14
WINDOWLESS, FLOW TYPE, PROPORTIONAL COUNTER FOR COUNTING C-14
复制标题
DOI:
10.1126/science.112.2903.198
复制
发表时间:
1950-01-01
期刊:
影响因子:
56.9
通讯作者:
ROBINSON, CV
中科院分区:
文献类型:
--
作者:
ROBINSON, CV
FIG. 1. Section of flow counter and sample changer. sulation, shielding, and line isolation of the circuits than is the case with Geiger-MUller counting. The entire counting apparatus consists of (1) the counter,(2) the sample changer,(3) the electronic unit, and (4) the gas supply tank and connections. The gas used is argon plus 5% CO, and is purchased already mixed. The gas flow is controlled by a needle valve with an extended handle and two stops, one for the flush position and one for steady flow. From the valve the gas is piped to an oil bubbler and thence to the counter. The gas can escape around the planchet and through the hole below.The electronic-unit is a Model 162 amplifier-scaler made by Nuclear Instrument and Development Labora-tories. The amplifier has a maximum gain of 400 and the scaler will pass. 40-v pulses. The counter is coupled directly into the amplifier input by a short piece of coaxial line. With this arrangement no preamplifier is required and the interference picked up in connections is minimized. An isolation transformer in the 110-v supply is necessary to prevent line interference. A section of the counter and the sample changer is shown in Fig. 1. The counter has a% &-in. diam brass cathode and an anode consisting of a loop of. 002-in.-diam tungsten wire. The loop is 9 mm by 17 mm and reaches to within 5 mm of the bottom of the planchet. The planchets used are of stainless steel and have a recess 9/16 in. in diam and. 037 in. deep. Measurements of the asymmetry of sensitivity of this counter have been made. For a small spot source placed on a planchet near the edge of the recess, therate of counting with the source in the plane of the loop is 10% more than at 900 from this position. For a reasonably uniform sample, there-fore, the asymmetry is negligible. The sample changer has a rotating disk as shown in Fig. 1, with two holes for planchets and a spring, not shown, which positions the planchets in the counter by catching in each of two notches in the rim of the disk.