对自掩模减影断层合成(SST)的图像质量和剂量节省进行了测试,SST将数字断层合成与模糊自掩模减影相结合。以中等剂量节省获得头部模型内耳的高质量图像。虽然他们采取了线性运动,他们没有表现出条纹由于偏离支点的对象,这是传统的线性断层扫描的特点。SST可以将内耳检查中的患者剂量降低至少12倍,并且机械方面可以通过对现有仪器进行适度修改来实现。
The image quality and dose savings of self-masking subtraction tomosynthesis (SST), which combines digital tomosynthesis with subtraction of a blurred self-mask, was tested. High-quality images of the inner ear of a head phantom were obtained at moderate dose savings. Although they were taken with linear motion, they did not exhibit the streaking due to off-fulcrum objects that is characteristic of conventional linear tomography. SST could reduce patient dose by a factor of at least 12 in examinations of the inner ear, and the mechanical aspects can be implemented with moderate modifications of existing instrumentation.