Comparison of Portapres non-invasive blood pressure measurement in the finger with intra-aortic pressure measurement during incremental bicycle exercise
Comparison of Portapres non-invasive blood pressure measurement in the finger with intra-aortic pressure measurement during incremental bicycle exercise
复制标题
DOI:
10.1097/00126097-200206000-00006
复制
发表时间:
2002-06-01
影响因子:
1.3
通讯作者:
Horstkotte, D
中科院分区:
文献类型:
--
作者:
Eckert, S;Horstkotte, D
Background and design In order to evaluate the accuracy of continuous, non-invasive blood pressure measurements in the finger during stress, blood pressure was measured in six patients with a Portapres Model 2 during increasing levels of bicycle exercise, using simultaneously registered intra-aortic (aortic arch) pressure as a reference.Methods Blood pressure was measured continuously and non-invasively in the left middle finger using the Portapres Model 2 device, at the same time invasively using a pigtail catheter situated in the aortic arch at rest and during and after standardized bicycle exercise. At the end of each stress level, in accordance with the protocol of the Association for the Advancement of Medical Intrumentation, systolic and diastolic blood pressure and pulse values were determined from the non-invasive (Portapres Model 2) and invasive measurements for 15 s each, and the mean values and standard deviations were calculated.Results Compared with simultaneous invasive blood pressure measurement in the aortic arch, the difference in measured blood pressure value (the value determined by the device minus the value determined invasively) at rest fell within the authorized normal range for German and European Wood pressure measurement devices: systolic +1.7 +/- 1.8 mmHg and diastolic -3.7 +/- 6.2 mmHg. The difference in measured value increased with the stress level.Conclusion Blood pressure and pulse can be determined reliably and accurately using the Portapres Model 2, non-invasively, both at rest and at low-to-moderate levels of stress.