10 kHz microsecond pulsed X-ray generator utilising a hot-cathode triode with variable durations for biomedical radiography
10 kHz microsecond pulsed X-ray generator utilising a hot-cathode triode with variable durations for biomedical radiography
复制标题
10 kHz 微秒脉冲 X 射线发生器,采用可变持续时间的热阴极三极管,用于生物医学放射成像
DOI:
10.1007/bf02512525
复制
发表时间:
1994
影响因子:
3.2
通讯作者:
T. Yanagisawa
中科院分区:
文献类型:
--
作者:
E. Sato;M. Sagae;Kei Takahashi;A. Shikoda;T. Oizumi;Y. Hayasi;Y. Tamakawa;T. Yanagisawa
A 10 kHz pulsed X-ray generator utilising a hot-cathode triode in conjunction with a new type of grid control device for controlling X-ray duration is described. The energy-storage condenser was charged up to 70 kV by a power supply, and the electric charges in the condenser were discharged to the X-ray tube repetitively by the grid control device. The maximum values of the grid voltage (negative value), the tube voltage, and the tube current were −1.5 kV, 70 kV, and 0.4 A, respectively. The duration of the flash X-ray pulse was primarily determined by the time constant of the grid control device and the cut-off voltage of thermoelectrons. The X-ray duration was controlled within a region of less than 1 ms; the X-ray intensity with a pulse width of 0.27 ms, a charged voltage of 70 kV, and a peak tube current of 0.4 A was 0.92 μC kg−1at 0.5 m per pulse. The maximum repetition rate was about 10 kHz, and the size of the focal spot was about 3.5×3.5 mm.