A directional fibre optic ultrasound transmitter based on a reduced graphene oxide and polydimethylsiloxane composite

A directional fibre optic ultrasound transmitter based on a reduced graphene oxide and polydimethylsiloxane composite
复制标题

DOI:
10.1063/1.5089750
复制
发表时间:
2019-03-18
影响因子:
4
通讯作者:
Desjardins, A. E.
Desjardins, A. E.
中科院分区:
物理与天体物理2区
文献类型:
--
作者:
Colchester, R. J.;Alles, E. J.;Desjardins, A. E.

文献摘要

被引文献

相似文献

Strongly directional ultrasound sources are desirable for many minimally invasive applications, as they enable high-quality imaging in the presence of positioning uncertainty. All-optical ultrasound is an emerging paradigm that exhibits high frequencies, large bandwidths, and a strong miniaturisation potential. Here, we report the design, modelling, and fabrication of a highly directional fibre-optic ultrasound transmitter that uses a composite of reduced graphene oxide and polydimethylsiloxane as the optical ultrasound generator. The ultrasound transmitter, which had an outer diameter of just 630 lm, generated ultrasound with a pressure exceeding 0.4 MPa for axial distances up to 16 mm, at a large bandwidth of 24.3 MHz. The ultrasound beam exhibited low divergence, with a beam diameter ranging between 0.6 and 2.1mm for distances between 0 and 16 mm. The presented directional optical ultrasound source is hence well-suited to high-resolution interventional imaging. (C) 2019 Author(s).