Preparation of calibrating standards for x-ray fluorescence spectrometry of trace metals in plastics
Preparation of calibrating standards for x-ray fluorescence spectrometry of trace metals in plastics
复制标题
塑料中痕量金属 X 射线荧光光谱测定校准标准品的制备
DOI:
10.1002/xrs.675
复制
发表时间:
2003
影响因子:
1.2
通讯作者:
Toshihiro Nakamura
中科院分区:
文献类型:
--
作者:
K. Nakano;Toshihiro Nakamura
Convenient x-ray fluorescent spectrometry was developed for the determination of V, Cr, Co, Ni, Ge and Sb in plastic materials. The calibrating standards were prepared by casting of polyurethane or polyester including a xylene solution of organometallic compounds as a standard. The calibrating disk had a homogeneous elemental distribution and excellent durability. The calibration curves of six elements showed a good linearity of less than 50 mass ppm for V and Ni, 100 mass ppm for Co and Ge, 200 mass ppm for Cr, and 400 mass ppm for Sb. The lower limits of detection (3σ) of the present method were 0.19 mass ppm for V, 0.42 for Cr, 0.24 for Co, 0.18 for Ni, 0.066 for Ge, and 2.1 for Sb. The proposed method has been applied to several different industrial plastics. The quantitative results are in good agreement with those obtained by the AAS method and x-ray fluorescent spectrometry using the fundamental parameter calculation method. Copyright © 2003 John Wiley & Sons, Ltd.