Characteristics of a CsI(Tl) Compton-suppressed clover detector
Characteristics of a CsI(Tl) Compton-suppressed clover detector
复制标题
DOI:
10.1088/0954-3899/40/3/035101
复制
发表时间:
2013-01
期刊:
影响因子:
--
通讯作者:
N. Zhang;Xiangguo Lei;Y. Guo;Xin Zhou;Y. Zhang;S. C. Li;Meilin Liu;Yufeng Zheng;Y. Fang;W. Hua;S. Guo;Y. Qiang;Jian-xiong Wang;B. Gao;S. W. Xu;S. Z. Chen
中科院分区:
文献类型:
--
作者:
N. Zhang;Xiangguo Lei;Y. Guo;Xin Zhou;Y. Zhang;S. C. Li;Meilin Liu;Yufeng Zheng;Y. Fang;W. Hua;S. Guo;Y. Qiang;Jian-xiong Wang;B. Gao;S. W. Xu;S. Z. Chen
The performance of a four fold segmented clover detector coupled to a CsI-suppressed shield has been tested using several standard radioactive sources. The summing characteristics, hit patterns and absolute efficiency were measured in both crystal and clover modes. Peak-to-total ratios of 57.3% and 45.6% in suppressed clover mode have been determined for 137Cs and 60Co sources, respectively. A Geant4 simulation of the suppressed clover detector, including the segmentation of the germanium crystals, has been created, and the simulation results agreed very well with the experimental data.