Construction, test and operation of a proton range radiography system
Construction, test and operation of a proton range radiography system
复制标题
DOI:
10.1016/j.nima.2010.11.096
复制
发表时间:
2011-02-11
影响因子:
1.4
通讯作者:
Watts, D.
中科院分区:
文献类型:
--
作者:
Amaldi, U.;Bianchi, A.;Watts, D.
We describe a Proton Range Radiography system with an imaging area of 10 x 10 cm(2), deploying a pair of Gas Electron Multiplier position-sensitive detectors and a scintillator stack to measure the residual proton range after crossing a target. The detector has been tested in the laboratory and in beam exposures at the Paul Scherrer Institute; images recorded with several phantoms of variable thickness and composition confirm the sub-millimetre accuracy and few percent energy resolution of the instrument. A new device with identical design and larger acceptance, 30 x 30 cm(2), is in construction. (C) 2010 Elsevier B.V. All rights reserved.