Reducing registration error in cross-beam vector doppler imaging with position sensor.
Reducing registration error in cross-beam vector doppler imaging with position sensor.
复制标题
使用位置传感器减少交叉束矢量多普勒成像中的配准误差。
DOI:
10.1109/iembs.2009.5332627
复制
发表时间:
2009
期刊:
影响因子:
--
通讯作者:
Kim,Yongmin
中科院分区:
文献类型:
--
作者:
Xu,Canxing;Beach,KirkW;Leotta,Daniel;Stuzman,Edward;Kim,Yongmin
Various vector Doppler methods have been proposed in the last several decades to overcome the Doppler angle dependency in both conventional spectral Doppler and color Doppler by measuring both the speed and direction of blood flow. However, they have not been adopted for routine use because most of them require specialized hardware, which is not available in commercial ultrasound systems. An alternative approach (cross-beam method) that uses color Doppler images obtained from different steered beam angles is more feasible, but there is error in registering multiple color Doppler images because they are not acquired simultaneously. To alleviate this problem, we have evaluated a cross-beam vector Doppler system that registers spatially with a position sensor two color Doppler images from two different angles and temporally with ECG synchronization. The registration error was reduced to an average of 0.92 mm from 2.49 mm in 9 human subjects. Vector Doppler carotid artery images of a healthy subject and a patient with atherosclerotic plaques are also presented.