An autonomous bladder pressure monitoring system
An autonomous bladder pressure monitoring system
复制标题
DOI:
10.1016/j.sna.2005.03.040
复制
发表时间:
2005-09-23
影响因子:
4.6
通讯作者:
Puers, R
中科院分区:
文献类型:
--
作者:
Coosemans, J;Puers, R
This paper reports on the development of an autonomous monitoring system, capable of continuously measuring the pressure inside the bladder. The capsule features wireless bi-directional communication and can be inductively powered anywhere inside the bladder. Short-circuiting the resonant LC tank for the data transmission maximizes the operating range of the passive telemetry. A novel clock extraction method is presented, based on lock-in of a Schmitt-trigger oscillator into the external radio-frequency field. A prototype was built using only discrete components. The system measures the pressure with a resolution of 0.04kPa at a sampling rate of 10 Hz. (C) 2005 Elsevier B.V. All rights reserved.