Method for improving the accuracy of continuous X-ray fluorescence analysis of iron ore mixtures
Method for improving the accuracy of continuous X-ray fluorescence analysis of iron ore mixtures
复制标题
DOI:
10.1134/s1061934810070129
复制
发表时间:
2010-07
影响因子:
1.1
通讯作者:
A. Volkov;N. Alov
中科院分区:
文献类型:
--
作者:
A. Volkov;N. Alov
The dependence of the intensity of X-ray fluorescence on the distance between the spectrometer and the test sample is studied. Changes in the intensity of primary X-radiation, absorption of X-rays by the air, and surface area of the analyzed material are calculated. An efficient method is proposed for improving the accuracy of continuous X-ray fluorescence analysis of iron ore mixtures on the conveyor belt through the exclusion of incorrect results by the dead time of the detection unit. The proposed method is universal and can be used for various loose materials.