High-sensitivity and wide-directivity ultrasound detection using high Q polymer microring resonators
High-sensitivity and wide-directivity ultrasound detection using high Q polymer microring resonators
复制标题
DOI:
10.1063/1.3589971
复制
发表时间:
2011-05-16
影响因子:
4
通讯作者:
Guo, L. Jay
中科院分区:
文献类型:
--
作者:
Ling, Tao;Chen, Sung-Liang;Guo, L. Jay
Small size ultrahigh Q polymer microrings working at near visible wavelength have been experimentally demonstrated as ultralow noise ultrasound detectors with wide directivity at high frequencies (> 20 MHz). By combining a resist reflow and a low bias continuous etching and passivation process in mold fabrication, imprinted polymer microrings with drastically improved sidewall smoothness were obtained. An ultralow noise-equivalent pressure of 21.4 Pa over 1-75 MHz range has been achieved using a fabricated detector of 60 mu m diameter. The device's wide acceptance angle with high sensitivity considerably benefits ultrasound-related imaging. (C) 2011 American Institute of Physics. [doi: 10.1063/1.3589971]