Comparison of the diagnostic accuracy of commercial NS1-based diagnostic tests for early dengue infection
Comparison of the diagnostic accuracy of commercial NS1-based diagnostic tests for early dengue infection
复制标题
DOI:
10.1186/1743-422x-7-361
复制
发表时间:
2010-12-06
期刊:
影响因子:
4.8
通讯作者:
Parra, Beatriz
中科院分区:
文献类型:
--
作者:
Osorio, Lyda;Ramirez, Meleny;Parra, Beatriz
Background: We compared the diagnostic accuracy and reproducibility of commercially available NS1-based dengue tests and explored factors influencing their sensitivities.Methods: Paired analysis of 310 samples previously characterized as positive (n = 218) and negative (n = 92) for viral isolation and/or RT-PCR and/or IgM seroconversion. Masked samples were tested by two observers with Platelia (TM) Dengue NS1 Ag, second generation Pan-E (TM) Dengue Early ELISA, SD Dengue NS1 Ag ELISA, Dengue NS1 Ag STRIP (TM), and SD BIOLINE (TM) Dengue Duo (NS1/IgM/IgG).Results: SD BIOLINE (TM) NS1/IgM/IgG had the highest sensitivity (80.7% 95% CI 75-85.7) with likelihood ratios of 7.4 (95% CI 4.1-13.8) and 0.21 (95% CI 0.16-0.28). The ELISA-format tests showed comparable sensitivities; all below 75%. STRIP (TM) and SD NS1 had even lower sensitivities (