Cerebellar transcranial magnetic stimulation: The role of coil type from distinct manufacturers.

Cerebellar transcranial magnetic stimulation: The role of coil type from distinct manufacturers.
复制标题

小脑经颅磁刺激:不同制造商的线圈类型的作用。

DOI:
10.1016/j.brs.2019.09.005
复制
发表时间:
2020
期刊:
影响因子:
7.7
通讯作者:
Spampinato D
Spampinato D
中科院分区:
医学1区
文献类型:
--
作者:
Spampinato D

文献摘要

相似文献

背景用经颅磁刺激刺激小脑通常被认为是不舒服的。目的研究Magstim(70 mm、110 mm涂层、110 mm涂层)、MagVenture和Deymed不同线圈对小脑刺激效果的影响。方法采用小脑-脑抑制实验方法,根据受试者的最大耐受刺激强度(MTI)绘制CBI招募曲线,以评估每个线圈激活小脑的效果。在MTI中,Magstim(110 mm涂层/非涂层)和Deymed线圈产生了可靠的CBI,而MagVenture线圈没有发现CBI。这些结果可以指导设计小脑TMS研究时线圈的选择和刺激参数。
BackgroundStimulating the cerebellum with transcranial magnetic stimulation is often perceived as uncomfortable. No study has systematically tested which coil design can effectively trigger a cerebellar response with the least discomfort.ObjectiveTo determine the relationship between perceived discomfort and effectiveness of cerebellar stimulation using different coils: MagStim (70 mm, 110 mm-coated, 110-uncoated), MagVenture and Deymed.MethodsUsing the cerebellar-brain inhibition (CBI) protocol, we conducted a CBI recruitment curve with respect to each participant’s maximum tolerated-stimulus intensity (MTI) to assess how effective each coil was at activating the cerebellum.ResultsOnly the Deymed double-cone coil elicited CBI at low intensities (−20% MTI). At the MTI, the MagStim (110 mm coated/uncoated) and Deymed coils produced reliable CBI, whereas no CBI was found with the MagVenture coil.Conclusions: The Deymed double-cone coil was most effective at cerebellar stimulation at tolerable intensities. These results can guide coil selection and stimulation parameters when designing cerebellar TMS studies.