Cardiac pacemakers and implantable cardioverter defibrillators:: In vitro magnetic resonance imaging evaluation at 1.5-Tesla
Cardiac pacemakers and implantable cardioverter defibrillators:: In vitro magnetic resonance imaging evaluation at 1.5-Tesla
复制标题
DOI:
10.1080/10976640600897237
复制
发表时间:
2007-01-01
影响因子:
6.4
通讯作者:
Fieno, David S.
中科院分区:
文献类型:
--
作者:
Shellock, Frank G.;Fischer, Laura;Fieno, David S.
Rationale and Objectives: To evaluate the effect of Magnetic Resonance Imaging (MRI) performed at 1.5-Tesla on current generation pacemakers and ICDs to identify safe parameters for MRI examinations. Methods: Pacemakers (Identity ADx XL DR+ 5386 and Identity ADx DR + 5380 generators; 1688T/52-cm atrial and ventricular leads) and ICDs (Atlas + V-243, Epic + V-236, and Epic + HF V-350 generators; Riata 1581/65-cm and QuickSite 1056K/75-cm leads; St. Jude Medical, Sylmar, California, USA) were evaluated for magnetic field interactions. MRI-related heating was assessed using various levels of RF power (SARs) and conditions that included scans on different body regions. Functional aspects of the devices were evaluated immediately before and after MRI procedures utilizing nine different pulse sequences. Induced currents were measured using a custom built system. Results: Magnetic field interactions will not create a hazard for these pacemakers and ICDs. All scans of the "head" and "lumbar" regions resulted in temperature changes