Improved reproducibility and quantitative analysis of phospholipids by flame ionization detection.
Improved reproducibility and quantitative analysis of phospholipids by flame ionization detection.
复制标题
通过氢火焰离子化检测提高了磷脂的重现性和定量分析。
DOI:
10.1016/0021-9673(85)80036-4
复制
发表时间:
1985
期刊:
影响因子:
--
通讯作者:
Murray,DK
中科院分区:
文献类型:
--
作者:
Murray,DK
Quantitative analysis of phospholipids by flame ionization was improved by careful application of samples with a Hamilton syringe and use of a sealed dual tank system. Chromarods developed more consistently with reproducible scanning times orRFvalues (coefficient of variation of 1%) and with sharper peaks if development was carried out in a sealed dual tank system. The Chromarods were placed in the inner tank, which was the standard ground glass topped tank furnished with the Iatroscan TH-10 system. This inner tank was placed inside a larger thin-layer chromatography tank which was sealed with silicone grease and the lid held in place with a lead brick. Both tanks were lined with absorbent paper and contained the same solvent system. Biological samples quantified with these procedures and measured in amounts between 1 and 30 μg had coefficients of variation between 0.2 and 6%. An efficient method of completely separating neutral lipids from phospholipids and allowing quantitative determination of cholesterol is desribed. Scanning times andRFvalues of various phospholipids are compared to determine the best separation of the major phospholipids found in 3T3-L1 and leukocyte membranes.