Simultaneous reconstruction of scintillation light and ionization charge produced by 511 keV photons in liquid xenon: Potential application to PET
Simultaneous reconstruction of scintillation light and ionization charge produced by 511 keV photons in liquid xenon: Potential application to PET
复制标题
DOI:
10.1016/j.nima.2009.06.036
复制
发表时间:
2009-08-21
影响因子:
1.4
通讯作者:
Sher, A.
中科院分区:
文献类型:
--
作者:
Amaudruz, P.;Bryman, D.;Sher, A.
In order to assess the performance of liquid xenon detectors for use in positron emission tomography we studied the scintillation light and ionization charge produced by 511 keV photons in a small prototype detector. Scintillation light was detected with large area avalanche photodiodes while ionization electrons were collected on an anode instrumented with low noise electronics after drifting up to 3 cm. Operational conditions were studied as a function of the electric field. Energy resolutions of