Double-Photon Emission Imaging With High-Resolution Si/CdTe Compton Cameras
Double-Photon Emission Imaging With High-Resolution Si/CdTe Compton Cameras
复制标题
DOI:
10.1109/tns.2021.3086799
复制
发表时间:
2021-08-01
影响因子:
1.8
通讯作者:
Shimazoe, Kenji
中科院分区:
文献类型:
--
作者:
Orita, Tadashi;Yabu, Goro;Shimazoe, Kenji
Application of Compton cameras to nuclear medicine or small animal imaging is a challenging subject. A major issue is insufficient spatial resolutions of the existing Compton camera systems. We developed an imaging system consisting of two Si/CdTe Compton cameras, which allows us to investigate gamma-gamma coincidence imaging for a variety of radioisotopes, including In-111, which emits double photons of 171 and 245 keV per decay and has been used for a long time in nuclear medicine and many commercialized radiotracers. We demonstrated with it a drastic reduction of a tail of the point spread function, compared with that of the traditional method, and obtained a spatial resolution of 4.5 mm full-width at half-maximum (FWHM) at 41.35 mm.