An ultrasmall angle scattering instrument for the DORIS‐III bypass
An ultrasmall angle scattering instrument for the DORIS‐III bypass
复制标题
用于 DORIS-III 旁路的超小角散射仪
DOI:
10.1063/1.1142730
复制
发表时间:
1992
影响因子:
1.6
通讯作者:
R. Gehrke
中科院分区:
文献类型:
--
作者:
R. Gehrke
The layout of the optics for a new small angle x‐ray scattering instrument at the DORIS‐III wiggler beamline BW4 was investigated by means of ray tracing and finite element calculations. The absolute intensity distributions at different locations of the beamline, especially in the focal plane, have been computed. The first crystal of a double crystal monochromator is the optical element that is most strongly influenced by thermal load. Thus, the distortion of this crystal surface due to thermal load was taken into account as well as the intended cooling system. The results show that the photon flux density at the focus is decreased by a factor of 0.2 due to this distortion.