Evaluation of image reconstruction according to changing physical parameter
Evaluation of image reconstruction according to changing physical parameter
复制标题
根据变化的物理参数评估图像重建
DOI:
10.1007/s10586-016-0546-0
复制
发表时间:
2016
期刊:
影响因子:
--
通讯作者:
Chang
中科院分区:
文献类型:
--
作者:
Min;Man;Jae;Yong;S. Seo;Gab;Chang
This study is to determine an optimum kernel, which objectively compares and analyzes the image according to physical parameters such as FBP, which is widely used in abdominal CT, as well as SAFIRE, which was developed recently. CT images were acquired with a 20-cm-diameter cylinder phantom filled with water and a SOMATOM Definition Edge CT (Siemens Healthcare). The images were reconstructed with a 3-mm-thick kernel such as FBP, which has B10 (very smooth), B20 (smooth), B30 (medium smooth), B40 (medium), B50 (medium sharp), and SAFIRE, which has I30 (strength 1 to 5) and I40 (strength 1 to 5). SNR, PSNR, RMSE, MAE and Plot profile were measured using the image J (version 1.49q; National Institutes of Health, Bethesda, MD, USA) in reconstructed images. In conclusion, the optimum kernel for abdominal CT should be a SAFIRE I30 kernel of strength 1.