Signal-to-motion artifact ratio versus frequency for impedance pneumography
Signal-to-motion artifact ratio versus frequency for impedance pneumography
复制标题
阻抗呼吸描记的信号与运动伪影比与频率的关系
DOI:
10.1109/10.364521
复制
发表时间:
1995
影响因子:
4.6
通讯作者:
J. Webster
中科院分区:
文献类型:
--
作者:
J. Rosell;J. Webster
The authors measured transthoracic impedance between 12.5 and 185 kHz in nine adults. They used a system with two impedance channels, both simultaneously detecting the real part of impedance at two different frequencies. They used only two electrodes in the midaxillary line, connecting both channels in parallel. The amplitude relation between the two channels was measured for different maneuvers and frequencies. Results show for normal breathing an increase of the signal of 20% and a decrease in motion artifacts from 12.5 to 185 kHz. It is concluded that, for the maneuvers studied, it is better to work at higher frequencies than the ones commonly used. Also, the authors suggest a method to further increase the signal-to-motion artifact ratio based on measurement at two frequencies.<<ETX>>