Assessing the limit of detection of Fourier-transform infrared spectroscopy and immunoassay strips for fentanyl in a real-world setting
Assessing the limit of detection of Fourier-transform infrared spectroscopy and immunoassay strips for fentanyl in a real-world setting
复制标题
DOI:
10.1111/dar.13004
复制
发表时间:
2019-11-19
影响因子:
3.8
通讯作者:
Ti, Lianping
中科院分区:
文献类型:
--
作者:
McCrae, Karen;Tobias, Samuel;Ti, Lianping
Introduction and Aims Drug checking is a harm reduction intervention increasingly used in the context of the opioid overdose epidemic. The aim of the study was to determine the limit of detection for fentanyl of two point-of-care drug checking technologies. Design and Methods Samples tested at point-of-care using Bruker Fourier transform infrared (FTIR) spectroscopy and BTNX fentanyl immunoassay strips were sent for confirmatory laboratory analysis using quantitative nuclear magnetic resonance (qNMR) spectroscopy. Concentrations by weight were determined and compared to results obtained with point-of-care methods. Results In total, 283 samples were sent for qNMR analysis; among these, 173 (61.1%) tested positive for fentanyl. As determined by qNMR, fentanyl concentration by weight ranged from 1% to 91%. Among these 173 samples, fentanyl was not detected in 30 (17.3%) samples by FTIR and in 4 (2.3%) samples by test strip. Samples containing fentanyl that went undetected by FTIR had concentrations