New Palpometer with Implications for Assessment of Deep Pain Sensitivity
New Palpometer with Implications for Assessment of Deep Pain Sensitivity
复制标题
DOI:
10.1177/0022034511402997
复制
发表时间:
2011-07-01
影响因子:
7.6
通讯作者:
Svensson, P.
中科院分区:
文献类型:
--
作者:
Futarmal, S.;Kothari, M.;Svensson, P.
The aims of this study were to compare test-retest variability and accuracy measures between (1) manual palpation and a novel palpometer and (2) different force levels. Sixteen clinicians were instructed to target 0.5, 1.0, and 2.0 kg on a force meter using manual palpation and the new palpometer (adjustable spring-coil with a small pin touching the examiner's hand when the correct pressure is reached). In all experiments, 10 consecutive measures of the force levels were recorded. The coefficient of variation (CV), actual force levels, and relative differences between target level and actual force level were compared between experiments and target levels. All outcome parameters had significantly lower values for the new palpometer than for manual palpation in a force-dependent manner (p < 0.004). CVs and actual force levels were significantly different between all target levels (p < 0.004). The new palpometer had low test-retest variability and provides a more accurate pressure stimulus than manual palpation.