Assessment of consistency in contouring of normal-tissue anatomic structures.
Assessment of consistency in contouring of normal-tissue anatomic structures.
复制标题
DOI:
10.1120/1.1521271
复制
发表时间:
2003-01-01
影响因子:
2.1
通讯作者:
Starkschall, George
中科院分区:
文献类型:
--
作者:
Collier, Dawn C;Burnett, Stuart S C;Starkschall, George
The purpose of this work is to estimate the uncertainty in the manual contouring of normal anatomical structures. The heart, esophagus, and spinal cord were contoured manually on six sets of computed tomography images by six dosimetrists whose experience ranged from 1 year to over 15 years. To determine the differences between inter- and intraobserver variations, each data set was contoured by one of the dosimetrists five times and once each by the five other dosimetrists. The magnitude of the discrepancies in delineating the contours was assessed. Intradosimetrist contouring discrepancies were as follows: esophagus, average 0.3 cm and maximum 2.9 cm; heart, average 0.5 cm and maximum 7.6 cm; and spinal cord, average 0.1 cm and maximum 0.7 cm. Interdosimetrist contouring discrepancies were as follows: esophagus, average 0.4 cm and maximum 3.1 cm; heart, average 0.7 cm and maximum 8.1 cm; and spinal cord, average 0.2 cm and maximum 0.9 cm. Significant discrepancies can occur when normal anatomic structures are contoured manually. Interdosimetrist discrepancies are typically slightly greater than intradosimetrist discrepancies. The magnitude of the discrepancies does not appear to be correlated to the experience of the dosimetrist.