Development of an enzymatic reaction device using magnetic bead-cluster handling
Development of an enzymatic reaction device using magnetic bead-cluster handling
复制标题
DOI:
10.1088/0960-1317/16/9/017
复制
发表时间:
2006-09-01
影响因子:
2.3
通讯作者:
Sato, Kazuo
中科院分区:
文献类型:
--
作者:
Shikida, Mitsuhiro;Takayanagi, Kentaro;Sato, Kazuo
We previously proposed a magnetic bead-cluster handling device for Micro-total analysis systems (Micro-TAS) and investigated its operation principle. The device does not need mechanical fluidic devices, such as pumps and valves, for handling solutions. We further developed the biochemical reaction unit chip, which is a key component in Micro-TAS, by applying a bead-cluster handling mechanism. We were able to do the enzymatic reaction by using hydrolysis between alkaline phosphatase and p-nitrophenyl phosphate. We also confirmed that the obtained calibration curves were linear during the enzymatic reaction. We had 70% reaction efficiency on the reaction chip by performing a comparative experiment. From these results, we concluded that our developed reaction chip is applicable to enzymatic immuno-assay systems.