A STANDARDIZED BREATH HOLDING TECHNIQUE FOR THE CLINICAL MEASUREMENT OF THE DIFFUSING CAPACITY OF THE LUNG FOR CARBON MONOXIDE
A STANDARDIZED BREATH HOLDING TECHNIQUE FOR THE CLINICAL MEASUREMENT OF THE DIFFUSING CAPACITY OF THE LUNG FOR CARBON MONOXIDE
复制标题
DOI:
10.1172/jci103402
复制
发表时间:
1957-01-01
影响因子:
15.9
通讯作者:
MORTON, JW
中科院分区:
文献类型:
--
作者:
OGILVIE, CM;FORSTER, RE;MORTON, JW
METHODSThe technique for the measurement of DL, which has been reported be'fore (1, 2), consists essentially of having the subject make a maximal inspiration of a gas mixture containing 10 per cent helium (He), 0.3 per cent CO and approximately 21 per cent 02 in N2 from the level of his residual volume, hold it for a measured time, and then rapidly expire. All of this expiration except the first liter is collected in a bag by the operator and analyzed as alveolar gas. The CO concentration which was present in this sample before any CO had been absorbed in the lungs is calculated from the dilution of the inspired He according to the equation,