AMS 14C Sample Preparation at the KCCAMS/UCI Facility: Status Report and Performance of Small Samples

AMS 14C Sample Preparation at the KCCAMS/UCI Facility: Status Report and Performance of Small Samples
复制标题

KCCAMS/UCI 设施的 AMS 14C 样品制备:小样品的状态报告和性能

DOI:
10.1017/s0033822200042181
复制
发表时间:
2007
期刊:
影响因子:
8.3
通讯作者:
Dachun Zhang
Dachun Zhang
中科院分区:
地球科学4区
文献类型:
--
作者:
Guaciara M. Santos;R. Moore;J. Southon;Sheila Griffin;E. Hinger;Dachun Zhang

文献摘要

被引文献

相似文献

我们提出了一个加速器质谱(AMS)放射性碳样品制备和测量的概述,描述了技术升级,现在使我们能够定期获得0.2-0.3%的精度为1毫克碳样品。在含100 μg碳的样品上也可以达到± 1%的精度。我们还开发了石墨化技术和AMS程序,用于超小样品(低至0.002 mg碳)。提供了大量和少量标准品的详细时间序列,如NIST OX-I和OX-II; FIRI-C和-D; IAEA-C6、-C7和-C8;以及不含14 C的样品。
We present an overview of accelerator mass spectrometry (AMS) radiocarbon sample preparation and measurements, describing the technical upgrades that now allow us to routinely obtain 0.2–0.3% precision for 1-mg carbon samples. A precision of ∼1% on samples with 100 μg of carbon can also be achieved. We have also developed graphitization techniques and AMS procedures for ultra-small samples (down to 0.002 mg of carbon). Detailed time series are presented for large and small aliquots of standards such as NIST OX-I and OX-II; FIRI-C and -D; IAEA-C6, -C7 and -C8; and 14C-free samples.