微少量環境試料測定用井戸型 Ge 検出器の効率特性
微少量環境試料測定用井戸型 Ge 検出器の効率特性
复制标题
井式Ge探测器测量少量环境样品的效率特性
DOI:
10.3769/radioisotopes.42.169
复制
发表时间:
1993
期刊:
影响因子:
--
通讯作者:
金井 豊
中科院分区:
文献类型:
--
作者:
金井 豊
A well-type Ge detector was studied to use for determining radioactivities of U and T h-series nuclides in small amount of environmental samples. Peak efficiencies of the well-type detector decreased against the height of sample (relative variation rate (a'/Eff (at24mm) ) was about -1%/ mm in this study) . The principal reason is thought to be in self-absorption. It is convenient to use the first-order regression formula to calculate the efficiencies if sample height is lower than the well depth. This technique was applied to lake core samples to determine sedimentation rate, and a good result was obtained.