文献摘要

被引文献

相似文献

本研究的目的是评估,在体外,陶瓷托槽和弓丝不同合金产生的摩擦力。对三种陶瓷托槽进行了性能测试:单晶(Aluminire ICE)、多晶(InVu)、带金属槽的多晶(Clarity)和一种不锈钢托槽(Dyna-Lock)。在人工唾液中,以0度和10度成角,对每个托槽的30个托槽进行测试,所有托槽都具有.022'槽,与不锈钢和镍钛丝.019' x.025 '组合。弓丝以10 mm/min的十字头速度穿过槽。研究的托槽组和弓丝组之间存在统计学显著差异(p <0.05)。带有金属槽的多晶托槽的值与传统多晶托槽的值相似,并且单晶托槽具有最高的摩擦力。镍钛丝产生的摩擦力最小。在多晶托槽中添加金属槽并没有显著降低摩擦值。镍钛丝产生的摩擦力低于不锈钢丝。
The aim of this study was to evaluate, in vitro, frictional forces produced by ceramic brackets and arch wires of different alloys. Frictional tests were performed on three ceramic brackets: monocrystalline (Inspire ICE), polycrystalline (InVu), polycrystalline with metal slot (Clarity), and one stainless steel bracket (Dyna-Lock). Thirty brackets of each were tested, all with .022' slots, in combination with stainless steel and nickel-titanium wires .019' x .025', at 0 degrees and 10 degrees angulation, in artificial saliva. Arch wires were pulled through the slots at a crosshead speed of 10 mm/min. There were statistically significant differences between the groups of brackets and wires studied (p < .05). The polycrystalline brackets with metal slots had values similar to those of conventional polycrystalline brackets, and the monocrystalline brackets had the highest frictional forces. The nickel-titanium wires produced the lowest friction. The addition of metal slots in the polycrystalline brackets did not significantly decrease frictional values. Nickel-titanium wires produced lower friction than those of stainless steel.