Making renal stones change size-impact of CT image post processing and reader variability
Making renal stones change size-impact of CT image post processing and reader variability
复制标题
DOI:
10.1007/s00330-011-2171-x
复制
发表时间:
2011-10-01
影响因子:
5.9
通讯作者:
Geijer, Hayenkan
中科院分区:
文献类型:
--
作者:
Liden, Mats;Andersson, Torbjoern;Geijer, Hayenkan
Objectives The objectives of this study were to quantify the impact of image post-processing parameters on the apparent renal stone size, and to quantify the intra- and inter-reader variability in renal stone size estimation.Methods Fifty CT datasets including a renal or ureteral stone were included retrospectively during a prospective inclusion period. Each of the CT datasets was post-processed in different ways regarding slice thickness, slice increment and window setting. In the first part of the study a single reader repeated size estimations for the renal stones using different post-processing parameters. In the intra-reader variability experiment one reader reported size estimations for the same images with a one-week interval. The inter-reader variability data were obtained from 11 readers reporting size estimations for the same renal stones.Results The apparent stone size differed according to image post-processing parameters with the largest mean differences seen with regard to the window settings experiment (1.5 mm, p < 0.001) and slice thickness (0.8 mm, p < 0.001). Changes in parameters introduced a bias and a pseudo-random variability. The inter-reader variability was considerably larger than the intra-reader variability.Conclusion Our results indicate a need for the standardisation of making measurements on CT images.