Liquid core waveguide-based optical spectrometry for field estimation of dissolved BTEX compounds in groundwater - A feasibility study

Liquid core waveguide-based optical spectrometry for field estimation of dissolved BTEX compounds in groundwater - A feasibility study
复制标题

DOI:
10.1016/s0003-2670(03)00423-9
复制
发表时间:
2003-06-06
影响因子:
6.2
通讯作者:
Dasgupta, PK
Dasgupta, PK
中科院分区:
化学1区
文献类型:
--
作者:
Larsson, H;Dasgupta, PK

文献摘要

被引文献

相似文献

检查了液芯波导 (LCW) 系统作为水中苯、甲苯、乙苯、二甲苯 (BTEX) 等化合物的哨兵监测器的实用性。制造了基于 Teflon AF 2400 管的适用于长光程吸收光谱的蒸汽渗透 LCW。使用基于光纤的闪光灯-光电二极管阵列 (PDA) 组合以及己烷作为波导芯中的溶剂进行近紫外多波长光谱分析。使用多组分校准,在 10 分钟的采样周期后,可以对混合物中的亚 mg l(-1) 水平的苯和甲苯进行定量,精确度可达 6+/-5%。 (C) 2003 Elsevier Science B.V. 保留所有权利。
The utility of a liquid core waveguide (LCW) system for acting as a sentry monitor for compounds such as benzene, toluene, ethylbenzene, xylene (BTEX) in water was examined. A vapor-permeable LCW suitable for long path length absorbance spectroscopy based on a Teflon AF 2400 tube was fabricated. Multiwavelength spectroscopy in the near-UV was carried out using a fiber optic-based flashlamp-photodiode array (PDA) combination with hexane as the solvent in the waveguide core. Using multicomponent calibration, quantitation of benzene and toluene accurate to 6+/-5% could be conducted at sub-mg l(-1) levels in mixtures after a sampling period of 10 min. (C) 2003 Elsevier Science B.V. All rights reserved.