Stereotactic accuracy and frame mounting: A phantom study.
Stereotactic accuracy and frame mounting: A phantom study.
复制标题
DOI:
10.25259/sni-88-2019
复制
发表时间:
2019-01-01
影响因子:
--
通讯作者:
Temel, Yasin
中科院分区:
文献类型:
--
作者:
Alptekin, Onur;Gubler, Felix S;Temel, Yasin
BACKGROUND: Frame mounting is considered one of the most critical steps in stereotactic neurosurgery. In routine clinical practice, the aim is to mount the frame as symmetrical as possible, parallel to Reid's line. However, sometimes, the frame is mounted asymmetrically often due to patient-related reasons.METHODS: In this study, we addressed the question whether an asymmetrically mounted frame influences the accuracy of stereotactic electrode implantation. A Citrullus lanatus was used for this study. After a magnetic resonance imaging scan, symmetric and asymmetric mounting of the frame, which could occur in clinical scenarios, was performed with computed tomography (CT). Three different stereotactic software packages were used to analyze the results. In addition, manual calculations were performed by two different observers.RESULTS: Our results show that an asymmetrically mounted frame (deviated, tilted, or rotated) does not affect the accuracy in the mediolateral axis (X-coordinate) or the anteroposterior axis (Y-coordinate). However, it can lead to a clinically relevant error in the superoinferior axis (Z-coordinate). This error was largest with manual calculations.CONCLUSION: These results suggest that asymmetrical frame mounting can lead to stereotactic inaccuracy in the superoinferior axis (Z coordinate).