Faraday-effect polarimeter-interferometer system for current density measurement on EAST.

Faraday-effect polarimeter-interferometer system for current density measurement on EAST.
复制标题

DOI:
10.1063/1.4889777
复制
发表时间:
2014-07
期刊:
The Review of scientific instruments
影响因子:
--
通讯作者:
H. Liu;Y. Jie;W. Ding;D. Brower;Z. Zou;W. Li;Zhongli Wang;J. Qian;Yanji Yang;L. Zeng
H. Liu;Y. Jie;W. Ding;D. Brower;Z. Zou;W. Li;Zhongli Wang;J. Qian;Yanji Yang;L. Zeng
中科院分区:
其他
文献类型:
--
作者:
H. Liu;Y. Jie;W. Ding;D. Brower;Z. Zou;W. Li;Zhongli Wang;J. Qian;Yanji Yang;L. Zeng

文献摘要

被引文献

相似文献

目前正在研制一种利用三波技术的多通道远红外激光偏振干涉仪(POINT)系统,用于EAST托卡马克装置的电流密度和电子密度分布测量。新型钼后向反射器安装在内壁上,用于双通光学装置。研制了一种用于POINT系统的250 kHz带宽的数字鉴相器,可实时提供法拉第旋转角和密度相移输出。初步标定表明,电子线积分密度分辨率小于5 × 10(16)m(-2)(± 2°),法拉第旋转角均方根相位噪声小于0.1°。
A multichannel far-infrared laser-based POlarimeter-INTerferometer (POINT) system utilizing the three-wave technique is under development for current density and electron density profile measurements in the EAST tokamak. Novel molybdenum retro-reflectors are mounted in the inside wall for the double-pass optical arrangement. A Digital Phase Detector with 250 kHz bandwidth, which will provide real-time Faraday rotation angle and density phase shift output, have been developed for use on the POINT system. Initial calibration indicates the electron line-integrated density resolution is less than 5 × 10(16) m(-2) (∼2°), and the Faraday rotation angle rms phase noise is <0.1°.