Enzyme Microelectrodes for Choline and Acetylcholine and Their Applications.
Enzyme Microelectrodes for Choline and Acetylcholine and Their Applications.
复制标题
胆碱和乙酰胆碱酶微电极及其应用。
DOI:
10.1016/0003-2670(94)00318-1
复制
发表时间:
1994
影响因子:
4.7
通讯作者:
Masashi
中科院分区:
文献类型:
--
作者:
Kano;Kenji;Morikage;Kazuo;Uno;Bunji;Esaka;Yukihiro;Masashi
Hydrogen peroxide based amperometric microsensors have been developed for choline (Ch) and acetylcholine (Ach) by coimmobilization of acetylcholinesterase (AchE) and choline oxidase (ChO) on platinum disk microelectrodes using glutaraldehyde vapor. The sensor was applied in micro flow injection and micro liquid chromatographic analyses. The chromatographic peak heights were linearly proportional to the amounts of Ach and Ch over the range of 0.05 to 103pmol (8 × 10−7to 1.5 × 10−2M, 0.06 μl) with correlation coefficients > 0.99. Immobilization of ChO was used for the selective determination of Ch with a response time of 5 s. This electrode was used for activity measurements of AchE down to 0.25 U ml−1in a 1 μl volume in a batch method at a total volume of 5.0 ml.
影响因子:
7.4
作者:
KRISTENSEN, EW;KUHR, WG;WIGHTMAN, RM
通讯作者:
WIGHTMAN, RM
影响因子:
7.4
作者:
BINDRA, DS;ZHANG, YN;REACH, G
通讯作者: