Inter-device Variability of the Stratus Optical Coherence Tomography

Inter-device Variability of the Stratus Optical Coherence Tomography
复制标题

DOI:
10.1016/j.ajo.2008.08.008
复制
发表时间:
2009-02-01
影响因子:
4.2
通讯作者:
Morad, Yair
Morad, Yair
中科院分区:
医学1区
文献类型:
--
作者:
Barkana, Yaniv;Burgansky-Eliash, Zvia;Morad, Yair

文献摘要

被引文献

相似文献

PURPOSE: To assess inter-device measurement variability with the Stratus optical coherence tomography (OCT).DESIGN: Evaluation of diagnostic test technology.METHODS: Eight eyes of eight healthy subjects were examined with four different Stratus machines in four medical centers during a period of five hours using fast retinal nerve fiber layer (RNFL) and fast macula protocols. Inter-device measurement variability and signal strength was assessed with standard deviation, coefficient of variance, and intraclass correlation coefficient (ICC). Spearman correlation coefficient was calculated between signal strength and thickness measurements.RESULTS: Statistically significant differences were not found for all macular parameters and all but one RNFL parameter. Mean signal strengths obtained with the four Stratus machines were significantly different; the newer the machine, the stronger signal strength it produced. Some RNFL parameters were moderately and statistically significantly correlated with signal strength.CONCLUSIONS: We did not observe statistically significant differences in any macular and all but one RNFL parameter between the four Stratus OCT machines. Signal strength was significantly different between the machines, and significantly correlated with some RNFL parameters. Inter-machine variability in RNFL measurements may have clinical significance. Signal strength should be included in the assessment of measurements obtained on different machines and in serial examinations with a single Stratus machine during patient follow-up. (Am J Ophthalmol 2009;147:260-266. (C) 2009 by Elsevier Inc. All rights reserved.)