A method for the analysis of low-mass molecules by MALDI-TOF mass spectrometry

A method for the analysis of low-mass molecules by MALDI-TOF mass spectrometry
复制标题

DOI:
10.1021/ac010979m
复制
发表时间:
2002-04-01
影响因子:
7.4
通讯作者:
Ni, JY
Ni, JY
中科院分区:
化学1区
文献类型:
--
作者:
Guo, Z;Zhang, QC;Ni, JY

文献摘要

被引文献

相似文献

我们报告了一种新的方法,称为基质抑制激光解吸/电离,以提高低质量分子的MALDI-TOF质谱分析。在该方法中,表面活性剂十六烷基三甲基溴化铵(CTAB)加入到α-氰基-4-羟基肉桂酸溶液的常规基质中,制备MALDI样品。在MALDI过程中,CTAB的存在可以基本上甚至完全抑制基质相关离子背景。因此,可以在低质量范围内常规获得非常干净的质谱。此外,CTAB的存在可以显著提高低质量分子的质量分辨率。可以看出,在基质/CTAB比为1000:1时常规获得高质量光谱。该方法已成功地用于分析各种低质量分子。
We report a novel method termed matrix suppressed laser desorption/ionization to improve the analysis of low-mass molecules by MALDI-TOF mass spectrometry. In this method, the surfactant of cetrimonium bromide (CTAB) is added to the conventional matrix of alpha-cyano-4-hydroxycinnamic acid solution to prepare the MALDI samples. During the MALDI process, the presence of CTAB could substantially or even completely suppress the matrix-related ion background. As a result, very clean mass spectra can be routinely obtained in the low-mass range. In addition, the presence of CTAB can significantly improve the mass resolution of low-mass molecules. It is seen that high-quality spectra were routinely obtained at a matrix/CTAB ratio of 1000:1. This method has been successfully used to analyze a variety of low-mass molecules.