Thin Film Nitinol Covered Stents: Design and Animal Testing
Thin Film Nitinol Covered Stents: Design and Animal Testing
复制标题
薄膜镍钛诺覆盖支架:设计和动物测试
DOI:
10.1097/mat.0b013e31816b43b0
复制
发表时间:
2008
期刊:
影响因子:
4.2
通讯作者:
G. Carman
中科院分区:
文献类型:
--
作者:
D. Levi;Ryan J Williams;Jasen Liu;Saar Danon;L. Stepan;M. Panduranga;M. Fishbein;G. Carman
Interventionalists in many specialties have the need for improved, low profile covered stents. Thin films of nitinol (<5–10 microns) could be used to improve current covered stent technology. A “hot target” sputter deposition technique was used to create thin films of nitinol for this study. Covered stents were created from commercially available balloon-inflatable and self-expanding stents. Stents were deployed in a laboratory flow loop and in four swine. Uncovered stent portions served as controls. Postmortem examinations were performed 2–6 weeks after implantation. In short-term testing, thin film nitinol covered stents deployed in the arterial circulation showed no intimal proliferation and were easily removed from the arterial wall postmortem. Scanning electron microscopy showed a thin layer of endothelial cells on the thin film, which covered the entire film by 3 weeks. By contrast, significant neointimal hyperplasia occurred on the luminal side of stents deployed in the venous circulation. Extremely low-profile covered stents can be manufactured using thin films of nitinol. Although long-term studies are needed, thin film nitinol may allow for the development of low-profile, nonthrombogenic covered stents.