NEUTRON RADIOGRAPHY AND DOSIMETRY IN HUMAN BEINGS - THEORETICAL STUDIES
NEUTRON RADIOGRAPHY AND DOSIMETRY IN HUMAN BEINGS - THEORETICAL STUDIES
复制标题
DOI:
10.1088/0031-9155/16/3/005
复制
发表时间:
1971-01-01
影响因子:
3.5
通讯作者:
PLECHATY, EF
中科院分区:
文献类型:
--
作者:
BUDINGER, TF;HOWERTON, RJ;PLECHATY, EF
Calculations were done for neutrons of 1 keV, 120 keV, fission spectrum, and 14 MeV, incident on a simulated human arm, to analyse the dose distributed to bone and marrow, and to determine the feasibility of neutron radiography as a practical clinical tool. The resolution possible through tissues as thick as 6 cm makes neutron radiography impractical except for very special circumstances. The dose calculations reveal as much as a twofold increase in dose deposited in the bone cortex compared with surrounding contiguous tissues. The skin dose is approximately twice the dose 3 cm inside the tissue. The average absorbed dose in the arm on the basis of a fluence of 10 10 neutrons incident per cm 2 is: 0.5, 1.6, 10, and 40.4 rads for 1 keV, 120 keV, fission spectrum and 14 MeV neutrons, respectively.