Geometrical properties of a digital beam attenuator system.
Geometrical properties of a digital beam attenuator system.
复制标题
数字光束衰减器系统的几何特性。
DOI:
10.1118/1.596086
复制
发表时间:
1987
期刊:
影响因子:
3.8
通讯作者:
Mistretta,CA
中科院分区:
文献类型:
--
作者:
Hasegawa,BH;Dobbins3rd,JT;Naimuddin,S;Peppler,WW;Mistretta,CA
A digital beam attenuator system has been developed to automatically generate patient‐specific compensating filters for chest radiography. An initial low‐dose test image is used to generate the attenuator, which is fabricated by overprinting multiple layers of a heavy‐metal material onto a nonattenuating substrate. The attenuator is subsequently inserted into the x‐ray beam for a final compensated radiograph. The effects of focal spot blurring and limited attenuator resolution result in the final compensated image containing only high‐spatial frequency information. The frequency response of the process is not strictly describable by a modulation transfer function, but an approximation of the frequencies remaining in the compensated image is obtained for low‐contrast conditions. It is found that a 4×4 blurring function on the original 64×64 test image is required for the attenuator to give appropriate compensated image appearance. A proposed attenuator printing scheme prints the attenuator in a 16×16 matrix, staggering successively printed layers to achieve the required 64×64 sampling with appropriate blurring. The resulting compensated image has good anatomical definition and contains a frequency response similar to that obtained by compensation techniques being investigated by Plewes and Sorenson.